BackgroundThere is limited data on diagnostic accuracy of recently introduced high-resolution Anger (HRA) SPECT incorporating attenuation correction (AC), noise reduction, and resolution recovery algorithms. We therefore studied 54 consecutive patients (excluding those with prior MI or cardiomyopathy) who had HRA-AC SPECT and coronary angiography (CA) ≤ 30 days and no change in symptoms. MethodsThe HRA-AC studies were acquired in 128 × 128 matrix (3.2 mm pixel) format with simultaneous Gd-153 line-source AC. Measured variables were image quality, interpretive certainty, sensitivity and specificity for any CAD, sensitivity for single- and multivessel CAD, and the influence of gender, body mass index (BMI), and stress modality. ResultsThe mean age of the patients was 66 ± 11 years with a BMI of 32 ± 7 kg·m−2. Mean interpretive certainty score was 2.7 on a 3-point scale and mean image quality score was 3.3 on a 4-point scale. Stress perfusion defects were detected in 34 of 38 patients with obstructive CAD [sensitivity 89%, 95% confidence interval (CI) 76%-95%]. The specificity was 75% (CI 51%-90%) and overall diagnostic accuracy was 85% (CI 73%-92%). Accuracy did not differ for females vs males, for BMI ≤30 vs >30, or for pharmacologic vs exercise SPECT. Sensitivity for single-vessel disease was 88% (CI 69%-96%) and for multivessel disease was 93% (CI 69%-99%). ConclusionNew Anger technology incorporating innovative improvements results in high image quality with excellent interpretive certainty and high diagnostic accuracy.