In situ magnetic hysteresis measurements of magnetic tips in a magnetic force microscope (MFM) are demonstrated using alternating gradient force magnetometry. The measured magnetic moments of MFM tips are estimated in the range from 10 −6 to 10 −5 emu by this technique and the whole MFM tips in cantilevers are considered to be measured from the value of measured magnetic moments. The relationship between the magnetic hysteresis loops of MFM tips and those of coated magnetic films is discussed.