To measure the degree of conversion (DC) of different 3D printing resins used for splints or orthodontic appliances under different postpolymerization conditions. Five 3D-printed photopolymer resins were studied. Each resin was analyzed in liquid form (n = 15), and then cylindrical specimens (n = 135) were additively manufactured and postcured with Form Cure (Formlabs) at different times (10, 60, and 90min) and temperatures (20°C, 60°C, and 80°C). The DC of each specimen was measured with Fourier transform infrared spectroscopy (FTIR). The data were statistically analyzed using a 3-way ANOVA followed by Tukey's post hoc test. The time and temperature of postpolymerization significantly influenced the DC of each resin: when time and/or temperature increased, the DC increased. For all resins tested, the lowest DC was obtained with a postcuring protocol at 10min and 20°C, and the highest DC was obtained at 90min and 80°C. However, at 80°C, the samples showed a yellowish color. With the Form Cure device, the time and temperature of postcuring could have an impact on the DC of the 3D printing resins studied. The DC of the 3D printing resins could be optimized by adjusting the postpolymerization protocol. Regardless of the resin used, when using the Form Cure device, postcuring at 60min and 60°C would be the minimal time and temperature conditions for achieving proper polymerization. Beyond that, it would be preferable to increase the postcuring time to boost the DC.