BackgroundThe high failure rates of metal on metal (MoM) large diameter head total hip arthroplasty (LDH THA) and hip resurfacing (HR) prevented their long-term comparisons with regards to clinical outcome. Such knowledge would be important as ceramic LDH bearing is now available. With long-term follow-up, we investigated the difference in 1) patient-reported outcome measures (PROMs); 2) revision and adverse events rates, and 3) metal ion levels between MoM LDH THA and HR.MethodsForty-eight patients were randomized for LDH THA (24) or HR (24) with the same MoM articulation. At a mean follow-up of 14 years, we compared between groups different PROMs, the number of revisions and adverse events, whole blood Cobalt (Co) and Chromium (Cr) ion levels, and radiographic signs of implant dysfunction.ResultsLDH THA (all cases: revised and well-functioning) had significantly better WOMAC (94 versus 85, p = 0.04), and more frequently reported having no limitation (p = 0.04). LDH THA revision rate was 20.8% (5/24) versus 8.3% (2/24) for HR (p = 0.4). Mean Co and Cr ion levels were higher in LDH THA compared to the HR (Co: 3.8 μg/L vs 1.7 μg/L; p = 0.04 and Cr: 1.9 μg/L vs 1.4 μg/L, p = 0.1). On radiographic analyses, 2 LDH THAs showed signs of adverse reaction to metal debris, whereas 1 loose femoral HR component was documented.ConclusionIn the long-term, MoM LDH THA had a high trunnion related revision rate but nonetheless showed better PROMs compared to HR. Provided with a well-functioning modular junction, non-MoM LDH THA would offer an appealing option.Trial registrationClinicalTrials.gov (NCT04516239), August 18, 2020. Retrospectively registered.