More than 2 million implantable collamer lenses (ICLs) have been implanted worldwide. With a central port to improve aqueous flow through the ICL, the latest iteration of this phakic intraocular lens (pIOL) has been shown to have stable outcomes with very low rates of adverse events. However, correct planning and ICL size selection continue to be important to achieve an optimal vault. Shallow or excessive vaults are not complications in and of themselves but may increase the risk of complications. Historically, surgeons have relied on measurements of anterior chamber depth (ACD) and manual, caliper-measured white-to-white (WTW) distance to select the ICL size. New diagnostic and imaging technologies such as optical coherence tomography (OCT) and ultrasound biomicroscopy (UBM) provide additional opportunities for visualization and measurement of the intraocular dimensions involved in phakic intraocular lens implantation, including sulcus-to-sulcus (STS) and angle-to-angle (ATA) diameters. This paper reviews various approaches to ICL planning and sizing that have been published in the peer-reviewed literature, all of which produce acceptable results for predicting vault and size selection. Surgeons may also want to identify a methodology for patient evaluation and ICL size selection that best aligns with their personal preferences, diagnostic technology, and familiarity with analytical optimization tools.
Read full abstract