Flow cytometry has recently established itself as a tool to track short-term dynamics in microbial community assembly and link those dynamics with ecological parameters. However, instrumental configurations of commercial cytometers and variability introduced through differential handling of the cells and instruments frequently cause data set variability at the single-cell level. This is especially pronounced with microorganisms, which are in the lower range of optical resolution. Although alignment beads are valuable to generally minimize instrumental noise and align overall machine settings, an artificial microbial cytometric mock community (mCMC) is mandatory for validating lab workflows and enabling comparison of data between experiments, thus representing a necessary reference standard for the reproducible cytometric characterization of microbial communities, especially in long-term studies. In this study, the mock community consisted of two Gram-positive and two Gram-negative bacterial strains, which can be assembled with respective subsets of cells, including spores, in any selected ratio or concentration. The preparation of the four strains takes a maximum of 5 d, and the stains are storable with either PFA/ethanol fixation at -20 °C or drying at 4 °C for at least 6 months. Starting from this stock, an mCMC can be assembled within 1 h. Fluorescence staining methods are presented and representatively applied with two high-resolution cell sorters and three benchtop flow cytometers. Benchmarked data sets allow the use of bioinformatic evaluation procedures to decode community behavior or convey qualified cell sorting decisions for subsequent high-resolution sequencing or proteomic routines.