Abstract

The more important properties of end-window Geiger counters, side-window proportional counters, and NaI(Tl) scintillation counters are reviewed and their performance in the 5 to 50 kvp x-ray analysis region is compared. Proportional and scintillation counters have resolving times of the order of 0.25 μsec and virtually eliminate the nonlinearity limitation of the Geiger counter. They produce pulse amplitudes proportional to the energy of the x-ray quanta, and thus simple pulse height discrimination techniques can be used to obtain greatly improved peak-to-background ratios with little loss of intensity of the characteristic x-rays. They are as reliable and nearly as easy to use as Geiger counters. The electronic discrimination method is described and explained with some practical applications. For a given experimental arrangement, the measured intensity is determined by the quantum counting efficiency of the counter tube, and the peak-to-background ratio is determined by the spectral response of the detector-discriminator combination.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.