Abstract

Highly integrated neural sensing microsystems are crucial to capture accurate signals for brain function investigations. In this paper, a 256-channel neural sensing microsystem with a sensing area of 5 × 5mm 2 is presented based on 2.5-D through-silicon-via (TSV) integration. This microsystem composes of dissolvable μ-needles, TSV-embedded μ-probes, 256-channel neural amplifiers, 11-bit area-power-efficient successive approximation register analog-to-digital converters, and serializers. This microsystem can detect 256 electrocorticography and local field potential signals within a small area of 5mm × 5mm. The neural amplifier realizes 57.8dB gain with only 9.8μW per channel. The overall power of this microsystem is only 3.79 mW for 256-channel neural sensing. A smaller microsystem with dimension of 6mm × 4mm has been also implanted into rat brain for somatosensory evoked potentials (SSEPs) recording by using contralateral and ipsilateral electrical stimuli with intensity from 0.2 to 1.0mA, and successfully observed different SSEPs from left somatosensory cortex of a rat.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.