Abstract

Since 2002, the postal audit in dental radiography has been supplementing standard quality control (QC) tools for dental intraoral X-ray machines. An aim of the audit is to check basic X-ray machine parameters (field size, exposure reproducibility), and a quality of the whole process of diagnostic imaging (entrance surface air-kerma measurement, a check of film processing and an image quality evaluation). The standard QC tests, performed by private companies, check mainly the X-ray unit. Conversely, the audit gives better information about the patient examination practices. During the period of January 2002 to May 2004 approximately 4000 audits were performed. The results confirmed that main problems in dental radiography are due to incorrect film processing, non-optimised setting of the exposure parameters and use of obsolete X-ray machines. Only approximately 30% of performed audits were satisfactory with respect to all checked parameters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.