Abstract
The aim of this study was to evaluate the occupational radiation exposure of staff during endoscopic retrograde cholangiopancreatography (ERCP), with a focus on individuals closest to the radiation source, and to identify potential increases in exposure to hands and eye lenses. Patient radiation exposure during ERCP was also assessed. Staff organ doses were monitored using a Philips Allura Xper FD 20 fluoroscopy system, during 24 ERCP procedures for a period of 7months. Staff doses were measured using thermoluminescence dosemeters and electronic personal dosemeters, and patient effective doses were simulated and calculated. Physicians' annual organ doses ranged from 0.2 to 1.6mSv for shoulders, 0.1 to 0.4mSv for eye lenses, and 0.3 to 1.6mSv for fingers. The annual organ dose of the nursing staff ranged from 0.08 to 2.4mSv for shoulders, 0.02 to 2.3mSv for eye lenses, and 1.2 to 5.3mSv for fingers. The effective dose to patients ranged from 0.009 to 0.46mSv. Staff doses were within safe limits, but patient doses were high, emphasizing the need for improved radiation protection.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.