Abstract

This study aims to estimate the entrance surface dose (ESD) of a water phantom for kilovoltage x-ray beams using an air kerma area product meter (KAP meter) equipped in an x-ray unit. The KAP meter was calibrated in terms of the ESD determined by a plane-parallel ionization chamber based on a 60Co absorbed dose-to-water calibration coefficient, ${N}_{D,w}^{{}^{60}\mathrm{C}\mathrm{o}}$. The ESD measured using the KAP meter was verified by comparing it with that estimated by the air kerma calibration coefficient, NK, for x-ray beam qualities. The ratio of ESDs based on ${N}_{D,w}^{{}^{60}\mathrm{C}\mathrm{o}}$ and NK was 1.003 on average and independent of the beam quality. The ESD by the KAP meter was an agreement within ±1.5% with that measured using the plane-parallel chamber for 10×10-30×30cm2 fields with a source-surface distance of 75-150cm. It was possible to estimate the ESD directly in a water phantom for x-ray beams without correction factors compared to the existing air kerma calibration, using a KAP meter calibrated based on ${N}_{D,w}^{{}^{60}\mathrm{C}\mathrm{o}}$.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call