Abstract
The International Commission on Radiological Protection recommends adaptation of the diagnostic reference levels (DRLs). Japan DRLs 2020 apply the entrance surface dose (ESD) in radiography. However, it is difficult to measure ESD in the clinical setting. A dose area product meter has been proposed for use as a dose index in interventional radiology. We investigated the basic characteristic of a dose area product meter and the relationship of ESD and dose area product meter values in radiography. We measured calibration factors from phantom studies and estimated ESD from the dose area product meter. Subject thickness was measured from the chest clinical images for calculation of ESD. Estimated ESD from the dose area product meter was compared with that calculated from program software (Surface Dose Evaluation Code, Sdec). Relative dose (dose area product meter/ionization chamber) decreased when tube voltage was lower (60 kV) or higher (130 kV). A positive correlation was found between the estimated and calculated ESD. Dose area product meter can be used for patient's dose control in radiography.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.