Abstract

Despite the growing use of dose management systems (DMSs), there is limited guidance on their optimal setup. This study aims to fill this gap by outlining the process followed to integrate "DOSE" (Qaelum NV, Belgium) DMS within the Cypriot public hospitals. DOSE was connected to the available central PACS, followed by a thorough data and functioning validation. A preliminary study was also conducted to assess the DOSE's ability to establish typical dose values in diagnostic radiography. Overall, over 70 x-ray units were configured. For some connections, it was necessary to first adjust the unit's DICOM attributes or to utilise the DICOM coercion technique. The DICOM image header was sometimes preferred over the DICOM RDSR, as the chosen data source. This study underscored the importance of understanding the equipment's technical capabilities for a successful DMS integration, while personnel with expertise in diagnostic radiology and the IT aspects of medical physics proved to have a vital role in the process. The pilot study showed that DOSE is an effective tool for establishing typical dose values and assessing compliance with relevant dose levels.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.