Abstract

The use of Ni-Cr thin-film resistors as strain-sensitive gages has been studied. Composition and thermal treatments are optimized for the following parameters: gage factor (FG) and temperature coefficient of resistance (TCR). TCR has been determined using an automatic data acquisition system with implemented software. FG has been determined by the cantilever method. The optimum thin-film composition (Ni 50%-Cr 50%) has been determined. Low stable TCR values (less than 20 ppm in the 30–150 °C temperature range) are obtained. Resistance stability is achieved with a 24 h heat treatment at 150 °C. The desired 40.Ω/square sheet resistance has been attained for 50 nm thick films. A longitudinal gage factor of 2.3 is obtained for this composition (and a lower transversal gage factor of 0.83). With the established process, Ni-Cr thin-film strain gages satisfy the main properties required for their application to piezoresistive pressure transducers: adequate strain sensitivity, high sheet resistance, low thermal sensitivity and good thermal stability.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.