Abstract

For monitoring of recombinant human antithrombin III during cell culture processes and subsequent purification steps a rapid method for quantitative determination was developed. The need for the introduction of this rapid method came from the limited availability of a quantitative enzyme-linked immunosorbent assay (ELISA) and the very time-consuming ELISA procedure. The developed method is based on reversed-phase high-performance liquid chromatography using a C4 column. The separation by gradient elution using water and acetonitrile takes less than 20 min even when complex samples, such as serum containing cell culture samples, have to be analyzed. Automation and a high sample throughput are possible with this reliable method. If necessary, insulin, transferrin and albumin can also be quantified with minor changes of the elution profile.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.