Abstract
2-D radiographs, while commonly used for evaluating sub-surface hard structures of teeth, have low sensitivity for early caries lesions, particularly those on tooth occlusal surfaces. Radiographs are also frequently refused by patients over safety concerns. Translucency of teeth in the near-infrared (NIR) range offers a non-ionizing and safe approach to detect dental caries. We report the construction of an NIR (850 nm) LED imaging system, comprised of an NIR source and an intraoral camera for rapid dental evaluations. The NIR system was used to image teeth of ten consenting human subjects and successfully detected secondary, amalgam–occluded and early caries lesions without supplementary image processing. The camera-wand system was also capable of revealing demineralized areas, deep and superficial cracks, and other clinical features of teeth usually visualized by X-rays. The NIR system’s clinical utility, simplistic design, low cost, and user friendliness make it an effective dental caries screening technology in conjunction or in place of radiographs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Journal of Translational Engineering in Health and Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.