Abstract

Among key design and operation issues for MgB2 relevant to MRI magnets are: uniformity of current-carrying capacity over long lengths (>2 km) of wire; and reliability of a splicing technique. This paper presents experimental results of current-carrying capacities of a small test coil and joints, both made from MgB2 round wires, multifilament and monofilament (mono), manufactured by Hyper Tech Research, Inc. The test coils were wound with 95-m long unreacted, C (carbon)-doped MgB2 multifilament wire, sintered at 700°C for 90 min. The critical currents were measured in the 4.2 K-15 K and 0 T-5 T ranges. We have modified our original splicing technique, proven successful with unreacted, un-doped MgB2 multifilament wire sintered at 570°C, and applied it to splice both un-doped and C-doped mono wires sintered at 700°C. Most consistently good results were obtained using the un-doped mono wires. Also presented are results of a small joint-coil-PCS assembly of mono wire, operated in persistent mode at 50 A at >10 K.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.