Abstract
We describe the conceptual design of a portable, liquid-helium-free, all-REBCO, 0.5-T/560-mm point-of-care magnetic resonance imaging (MRI) magnet. It is free from an external power supply and a refrigeration system during operation. In our portable MRI magnet, we use a detachable "cryocirculator" that circulates, in a closed circuit, cold working fluid, and most importantly for portability, it can be readily coupled to or decoupled from the magnet, in contrast, a conventional cryocooler is mechanically attached to the magnet. Another unique feature of our system is a volume of solid nitrogen (SN2) in the cold chamber that adds enough thermal mass to the magnet in the 30-36-K operating temperature range, enabling it to maintain its field over a period of, for this system, ≥10 hours, plenty enough for this portable MRI system, uncoupled from its cryocirculator, to perform its mission before it needs recooling.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE transactions on applied superconductivity : a publication of the IEEE Superconductivity Committee
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.