Abstract

Bioanalysts are continuously looking for innovative ideas or instruments to increase the sensitivity and selectivity of their assays. Research for better mass spectrometers is becoming crucial with the emerging trend of large-molecule quantification. This study lists the different advantages of high-resolution MS (HRMS) over standard triple quadrupole instruments and proposes basic guidelines on how to use HRMS for large-molecule quantification in a regulated environment. A direct comparison between HRMS and triple quadrupole instruments for the quantification of six different model peptides (desmopressin, calcitonin, enfuvirtide, exenatide, glucagon and somatostatin) was completed. The HRMS instrument, when used specifically for targeted quantification ('quant/quant'), showed equivalent or better sensitivity for all compounds tested. This paper demonstrates that the use of a HRMS instrument in a regulated environment is a viable technique for quantification of large molecules. The latter was able to allow flexibility and selectivity to adapt the specificity of each assay with sensitivity comparable to the triple quadrupole instrument.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.