Abstract

AbstractThe influence of the crossover effect on the tube voltage dependence of the quantum mottle for the front and back emulsion has been investigated. A new method is proposed for separating the Wiener spectra of the screen mottle for the front and back emulsions with the crossover effect into those of the quantum mottle and the structure mottle and a method for obtaining the detective quantum efficiency (DQE) and the noise-equivalent number of quanta (NEQ).The numbers of incident X-ray photons with the crossover effect were smaller than those without the effect for all tube voltages studied. However, at tube voltages below about 75 or 80 kV, the Wiener spectral values of the quantum mottle with the crossover effect at 0 mm 1 were smaller than those without the effect. The Wiener spectral value increased with the tube voltage and, at tube voltages above about 75 or 80 kV, exceeded that without the effect. Because the Wiener spectrum of the quantum mottle is proportional to the reciprocal of the NEQ, th...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.