Abstract

Flow injection and reversed flow injection analysis with amperometric detection were used for the determination of the concentrations of strong acids and strong bases and the pH of the solutions. The amperometric detection was based on the reduction current of hydrogen ion or water molecules. A platinum indicator electrode cathodically polarized at a constant potential was employed, choosing the potential in the rising part of the current/potential curve. On the basis of the experimental data an equation was developed for the calculation of the concentrations of the strong acids and strong bases and the pH by a non-linear regression computer program.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.