The combination of flow injection technique and spectrophotometric detection can be employed for the rapid and low-cost methods for antibiotic assay. In this work, two flow injection analysis (FIA) systems, normal and reverse (nFIA and rFIA), combined with UV-Vis spectrophotometric technique were used for amoxicillin determination in its pure and pharmaceutical preparations. A colorimetric coupling reaction between amoxicillin and diazotized p-toluidine produced a bright yellow azo dye and quantified at a maximum wavelength of 426 nm. For nFIA and rFIA procedures, the amoxicillin calibration graphs had an RSD of less than 2% and were linear in the ranges of 5-200 and 1-140 µg/mL, respectively. The limits of detection and quantification were 1.41 and 4.71 µg/mL of amoxicillin for nFIA and 0.39 and 1.31 µg/mL for rFIA, respectively. The physical and chemical factors that could influence color sensitivity were investigated. Various commercial formulations containing different amounts of amoxicillin were successfully tested using the developed approaches, which demonstrated to be easy, rapid, reproducible, and interference-free.
 KEY WORDS: Amoxicillin, Normal FIA, Reverse FIA, p-Toluidine, Diazotization reaction
 Bull. Chem. Soc. Ethiop. 2024, 38(3), 577-590. 
 DOI: https://dx.doi.org/10.4314/bcse.v38i3.3

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.