Abstract

Abstract In this work, we describe for the first time the use of the internal standard method in flow injection analysis (FIA) with amperometric detection. The method is based on the application of sequential potential pulses to the working electrode in an electrochemical flow cell. The sequence of potential pulses is selected in such a way that the analyte and internal standard compound are detected and monitored individually and independently at the same working electrode. This approach compensates for random errors associated with variations of flow rate, injection volume, ionic strength difference between standards and samples, and accidental insertion or formation of air bubbles in the carrier stream. In addition, this method can overcome the major drawback of amperometric detection using solid electrodes, which is gradual electrode passivation. To illustrate the potential of this method, the flow-injection amperometric detection of uric acid using [Fe(CN) 6 ] 3− as an internal standard (IS) is presented as an example.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.