Abstract

In radiographs, the Wiener spectral values al low frequency have been used for the evaluation of radiographic mottle. The validity of the psychophysical RMS granularity a,, as a new measure for the evaluation of radiographic mottle was confirmed in our previous paper IJ. Photogr. Sci., 43, 103 (1995)1.We calculate the density dependence at various frequencies of the Wiener spectrum W(v) and the psychophysical Wiener spectrum Wp(v), which is psycliophysically corrected, and compare with that of σp2. The density dependence of Wr(v) at a frequency of 2.0 mm -1 is in good agreement with that of σp2. It is therefore possible to evaluate the radiographic mottle using the W/v) value, at a frequency of 2.0 mm -1 instead of ap in this example. The time to obtain the density dependence of ap, which requires the measurement of ΔD,mm, values at many frequencies and integration with respect to frequency, can be reduced using this method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.