Abstract

A compact MRI system for trabecular bone (TB) microstructure measurements of the distal radius was developed using a 1.0 T permanent magnet and a compact MRI console. TB microstructure of the distal radius was clearly visualized using a three-dimensional (3D) driven equilibrium spin-echo (DESE) sequence in 23 min. The image obtained had a sufficient spatial resolution (150 microm x 150 microm x 500 microm) and signal-to-noise ratio (SNR) (approximately 10) for 3D bone microstructure analysis. The system demonstrated the feasibility of using a permanent magnet compact MRI system as a clinical instrument for bone microstructure measurements of the distal radius.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.