Abstract

A compact MRI system for measuring the trabecular bone (TB) microstructure of the finger using a high-field-strength (1.0T) permanent magnet was developed. The entire system was installed in a 0.6 mx1.2 m space. One male and 36 female subjects participated in the imaging experiments. The TB of the distal phalanx of the middle finger was imaged at a voxel resolution of (160 microm)3 using a three-dimensional (3D) driven equilibrium spin-echo (SE) imaging sequence (imaging time=approximately 14 min). The image data sets obtained yielded two distinct peaks for the bone and marrow in image intensity histograms when no motion was present. The structural parameters obtained through 3D image analysis show that this compact system is potentially useful for evaluating bone quality.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.