Abstract

A rapid and selective HPLC-MS/MS method was developed for the determination of LBPT in human plasma. The analyte was extracted from plasma samples by solid-phase extraction and then chromatographed on a C18 analytical column. The mobile phase consisted of acetonitrile-10mM ammonium formate in 0.1% formic acid (30:70, v/v) and the flow rate was 0.2 mL/min. The detection was performed on a triple quadrupole tandem mass spectrometer in multiple reactions monitoring (MRM) mode using positive electrospray ionization (ESI). The method was validated over the concentration range of 0.2-100 ng/mL. Inter- and intra-day precision (RSD %) were less than 9.2% and the accuracy (RE %) ranged from 0 to 11.0%. The lower limit of quantitation (LLOQ) was 0.2 ng/mL. The extraction recovery was on average 75% and the detection was not affected by the matrix. The method was successfully applied to the pharmacokinetic study of LBPT in healthy Chinese subjects.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.