Abstract

The goal of work is the characterization of two PET/CT units that are based on different technologies, using the methodology proposed in the NEMA NU 2-2001 standard. The two systems were qualified in the 3D acquisition mode by means of the National Electric Manufacturers Association (NEMA) phantoms. The results obtained showed that the NEMA standard allows to highlight differences in terms of spatial resolution, sensitivity, scatter fraction and counting rate performances between the two systems ; differences that can be explained by the geometry of the units and the materials of the detectors used. Thus, the use of the standard makes it possible to benchmark PET units and establish reference values that can be used to follow the stability of the system over the time. The tests of image quality, obtained under conditions closer to the clinical applications, showed nevertheless that the two units give, in spite of different technologies, images with sufficient contrast, within short acquisition times, allowing the detection of small lesions. This test should be part of the constancy tests to enable the comparison of examinations performed on different units when the quantitative aspects are of importance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.