Abstract

A crystal-optic x-ray fluorescence energy analyzer has been designed and tested, which combines the features of electron-volt energy resolution, large solid angle coverage, and tunability over several kilo-electron-volts. The design is based upon the principle of active optics, with ten actuators available to optimally adjust the shape of a silicon crystal used in the Bragg geometry. In most applications the shape is that of a logarithmic spiral for high energy resolution with a spatially nonresolving detector, but a wide range of other shapes is also possible for applications such as imaging or single-shot spectroscopy in a spectral range of the operator's choosing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.