Abstract

An experiment measuring the 2S Lamb shift in muonic hydrogen (/spl mu//sup -/p) is being performed at the Paul Scherrer Institute (PSI), Switzerland. It requires small detectors for 1.9 keV X-rays (2P-1S transition) with an energy resolution around 25%, a time resolution better than 100 ns, a large solid angle coverage, and insensitivity to a 5 T magnetic field. We used large area avalanche photodiodes (LAAPDs) from radiation monitoring devices (RMD) for the latest data taking period in 2003. For soft X-ray spectroscopy applications, they have to be cooled, hence, a series of tests were performed to choose the optimal operation temperature. The temperature dependence of gain, energy resolution, dark current, excess noise factor, and detector response linearity was studied. Finally, details of the LAAPDs application in our experiment as well as their response to alpha particles are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.