Abstract
The design and operating characteristics of a fully differential pressure transducer are described. The device is intended for use with He II heat transfer experiments where it operates in vacuum and at low temperatures (T<4.2 K). A movable electrode is attached to two sets of miniature bellows such that the electrode position is determined by the differential pressure across the device. The movable electrode is located between two fixed electrodes, thus forming a pair of variable capacitors. A dedicated charge amplifier is used to convert the pressure induced capacitance change to an ac output voltage. The sensitivity is roughly 5 μV/Pa. For the present application, the capacitor and electronics have acceptable performance, with a mean noise level of ±5 Pa.