A modified electrode configuration in a spark source mass spectrometer improves the reproducibility of analytical results. The system consists of vertically mounted cylindrical electrodes facing each other, one of which is rotating. No rotary feedthrough is needed as coupling between a small magnet inside the source and one (driven by a motor) outside allows rotation. Hence the vacuum inside the source can be kept low (less than 10-6 Pa approximately=10-8 Torr). Relative standard deviations do not exceed 6% for highly inhomogeneous powder mixtures. Moreover, higher and more stable transmissions are obtained, while short-circuiting between low-melting point materials is greatly reduced.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.