Purpose: In the hyperthermia treatment of deep-seated tumours by a phased array of radiofrequency (RF) antennas, heatability will be investigated in terms of power-to-tumour and other figures-of-merit of hyperthermia treatments to be optimised. The assumption is that each source is individually constrained to not exceed a maximal nominal power. The nominal power may differ from a source to another as a physical limit or an operative modality.Method: Under such constraint, new procedures for the maximisation of (i) power-to-tumour, (ii) heating efficiency and, in general, (iii) power ratios as tumour-heating selectivity are proposed. (iv) The problem whether a tumour is equally heatable after turning off some antennas is addressed as array thinning.Case study: An array of eight dipoles arranged on two lines around a head/neck is introduced to perform a numerical analysis. The achievable power-to-tumour according to the new optimizations and other performance indices adopted from the literature is tested against values of power that can be found to be sufficient for heating tumours to clinical temperatures. New solutions to data rendering in hyperthermia heating are proposed.