We present a systematic method to absolutely calibrate detector efficiency vs photon energy using a laser produced plasma broadband x-ray source, a gold standard calibrated detector, and transmission gratings (TGs) as dispersive elements. Calibration uses one calibrated TG and a calibrated gold standard detector on one channel and a second calibrated TG and a detector to be calibrated on the other channel. Both channels simultaneously view the laser-produced plasma x-ray source from the same angle with respect to the laser beam and the planar target normal. Image plate detectors are calibrated for the first time at photon energies below 700 eV. Single shot simultaneous calibration of several detectors is possible, making this method an efficient and practical way to periodically calibrate detectors, using in-house capabilities of laser laboratories.