Abstract

We present a systematic method to absolutely calibrate detector efficiency vs photon energy using a laser produced plasma broadband x-ray source, a gold standard calibrated detector, and transmission gratings (TGs) as dispersive elements. Calibration uses one calibrated TG and a calibrated gold standard detector on one channel and a second calibrated TG and a detector to be calibrated on the other channel. Both channels simultaneously view the laser-produced plasma x-ray source from the same angle with respect to the laser beam and the planar target normal. Image plate detectors are calibrated for the first time at photon energies below 700 eV. Single shot simultaneous calibration of several detectors is possible, making this method an efficient and practical way to periodically calibrate detectors, using in-house capabilities of laser laboratories.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.