Purpose. Accuracy of dose delivery in multiple breath-hold segmented volumetric modulated arc therapy (VMAT) was evaluated in comparison to noninterrupted VMAT using a static phantom. Material and Methods. Five VMAT plans were evaluated. A Synergy linear accelerator (Elekta AB, Stockholm, Sweden) was employed. A VMAT delivery sequence was divided into multiple segments according to each of the predefined breath-hold periods (10, 15, 20, 30, and 40 seconds). The segmented VMAT delivery was compared to noninterrupted VMAT delivery in terms of the isocenter dose and pass rates of a dose difference of 1% with a dose threshold of 10% of the maximum dose on a central coronal plane using a two-dimensional dosimeter, MatriXX Evolution (IBA Dosimetry, Schwarzenbruck, Germany). Results. Means of the isocenter dose differences were 0.5%, 0.2%, 0.2%, 0.0%, and 0.0% for the beam-on-times between interrupts of 10, 15, 20, 30, and 40 seconds, respectively. Means of the pass rates were 85%, 99.9%, 100%, 100%, and 100% in the same order as the above. Conclusion. Our static phantom study indicated that the multiple breath-hold segmented VMAT maintains stable and accurate dose delivery when the beam-on-time between interrupts is 15 seconds or greater.