Introduced here is the on-line coupling of hollow-fiber flow field-flow fractionation (HF5) to depolarized multi-angle static light scattering (D-MALS). HF5 is a size-based separation alternative to size-exclusion and hydrodynamic chromatography and asymmetric flow field-flow fractionation. HF5 can separate larger sizes than its chromatographic counterparts and provides several advantages over its fractionation counterpart, including reduced sample consumption and greater ease of operation. D-MALS is a variant of MALS in which the depolarized scattering from the analyte solution is measured at a variety of angles simultaneously. Measurements of depolarized scattering have previously been employed in studying the optical properties of solutions or suspensions, to determine the length of rod-like analytes, and to gain increased accuracy in the determination of analyte molar mass. The coupling HF5/D-MALS allows for the depolarization ratio of a solution or suspension to be measured continuously across the fractogram. This is demonstrated here for a Teflon latex the size range of which extends beyond that accessible to commercial size-exclusion columns. The results presented provide the first reported on-line HF5/D-MALS coupling, showing the feasibility of the technique as well as its realized potential for providing continuous depolarization measurements, inter alia.