Introduction: The Nucleus Straight Research Array (SRA) cochlear implant has a new 25-mm electrode carrier designed to minimize insertion trauma, in particular allowing easy insertion via the round window. The aims of this study were to measure preoperative to postoperative benefit in terms of speech recognition in quiet and in noise in three groups of patients (electrical complement, EC; electrical stimulation, ES; electro-acoustic stimulation, EAS) with varying levels of low-frequency hearing, and to evaluate the preservation of residual hearing after implantation with the SRA cochlear implant. Methods: The study design was prospective with sequential enrolment and within-subject comparisons: 23 adult cochlear implant candidates were divided into three groups according to their level of preoperative residual hearing at 500 Hz (EC ≤50 dB; 50 dB < EAS < 80 dB; ES ≧80 dB). Monosyllabic word recognition using the SRA cochlear implant in combination with residual low-frequency hearing was assessed at 4 and 13 months after implantation. Hearing threshold levels were also monitored over time. Results: Subjects across all three groups had significant improvements in speech recognition scores (i.e. >20 percentage points) both for listening in quiet (71% of subjects) and in noise (100% of subjects). The average score at 4 months after operation for words presented in quiet was 61.7%, and in 10 dB SNR noise 46.5%, compared to 34.4 and 10.6% preoperatively (p < 0.001). All subjects retained measurable hearing at 500 Hz in the implanted ear at 4 months after the operation; mean increases were 19, 29 and 1 dB for the EC, EAS and ES groups (n = 21). Across frequencies of 125–1000 Hz, the median increase in thresholds was 15 dB up to 13 months postoperatively (n = 15). Conclusions: Speech recognition performance of subjects with various levels of residual low-frequency hearing was significantly improved with the SRA cochlear implant. A high level and rate of hearing preservation was achieved with the SRA implanted using a round window surgical technique. Subjects with preoperative low-frequency hearing levels between 50 and 80 dB HL (EAS group) tended to lose more hearing than those with either better or worse hearing.
Read full abstract