Quadrupole mass spectrometers are widely used, and voltage scanning is their traditional working mode. By fixing the scanning voltage frequency and changing the value of the RF voltage, ions with different mass numbers can reach the detector in sequence, achieving ion selection. When analyzing high-mass molecules, several kilovolts of scanning voltage are required, which is not conducive to the miniaturization and safety of the instrument. By selecting the frequency of the scanning RF power supply and fixing the value of the RF power supply voltage, ion selection can be achieved by changing the frequency of the RF power supply, enabling miniaturized mass spectrometry analysis of high-mass molecules. In this paper, a high-speed gallium nitride driver circuit for frequency scanning mass spectrometry analysis is designed. The NCP51820 high-speed gate driver and INN650D140A gallium nitride MOS tube are selected to form a full-bridge driver, realizing a quadrupole rectangular wave RF power supply. The system has a maximum withstand voltage of 650 V and a frequency range of 400 K–4 MHz, allowing for scanning measurements of mass numbers ranging from 3 to 606 amu.