Background: The purpose of this study was to evaluate influence of technical parameters on Stereotactic Radiosurgery (SRS) plans on dosimetric evaluation. Materials and methods: There were considered and selected two patients who had brain tumors and had been treating in Tomsk Regional Oncology Center. All treatment plans were simulated using the Monaco treatment planning system v 5.11 (Elekta Instrument AB, Stockholm) on the Elekta Synergy linac with photon beam nominal energy 6 MV. For study there were created several plans with different technical parameters (number of arcs, increment, gantry angle, collimator angle, couch angle, minimum segment width). Each plan was evaluated by metrics such as Conformal, Homogeneity and Gradient indexes. Results: If we use a large increment value it may create few sectors and which can produce low quality plans, and increase treatment time, otherwise if we use a too small increment value it will give us more sectors and they can increase the quality of the plan. The dose which was measured with higher MSW showed to us better agreement with the calculated dose. Conclusion: It is valuable to carefully select technical parameters. The use of rigid increment value, segment width or increasing number of arcs can make plans have lower quality and to rise time of treatment.