Clinical trials have demonstrated the efficacy and safety of hybrid closed-loop (HCL) systems, yet few studies have compared outcomes in the real-world setting. This retrospective study analyzed patients from an academic endocrinology practice between January 1, 2018, and November 18, 2022. The inclusion criteria were diagnosis code for type I diabetes (T1D), >18 years of age, new to any HCL system [Medtronic 670G/770G (MT), Tandem Control IQ (CIQ), or Omnipod 5 (OP5)], and availability of a pump download within three months. The outcomes included %time in range (TIR) of 70 to 180 mg/dL, %time below range (TBR) <70 mg/dL at 90 days, and HbA1c for 91 to 180 days. Of the 176 participants, 47 were MT, 74 CIQ, and 55 OP5. Median (25%, 75%) change in HbA1c was -0.1 (-0.8, 0.3), -0.6 (-1.1, -0.15), and -0.55 (-0.98, 0)% for MT, CIQ, and OP5, respectively, (P = .04). TIR was 70 (57, 76), 67 (59, 75), and 68 (60, 76)% (P = .95) at 90 days while TBR was 2 (1, 3), 1 (0, 2), and 1 (0, 1)%, respectively, (P = .002). The %time in automated delivery was associated with TIR and change in HbA1c. After controlling other factors including %time in automated delivery, HCL type was not an independent predictor of change in HbA1c nor TIR but remained a significant predictor of TBR. There were significant reductions in HbA1c in CIQ and OP5. TIR was similar across pumps, but TBR was highest with MT. The %time in automated delivery likely explains differences in change in HbA1c but not TBR between HCL systems.