Abstract

The aim of this study was to assess the efficacy and safety of hybrid closed-loop (HCL) systems for insulin delivery in children and adolescents with type 1 diabetes (T1D). We searched Embase, PubMed, and Cochrane Library for randomized controlled trials (RCTs) published until March 2023 comparing the HCL therapy with control therapies for children and adolescents with T1D. We computed weighted mean differences (WMDs) for continuous outcomes and risk ratios (RRs) with 95% confidence intervals (CIs) for binary endpoints. Four RCTs and 501 patients were included, of whom 323 were randomized to HCL therapy. Compared with control therapies, HCL significantly improved the period during which glucose level was 70-180 mg/dL (WMD 10.89%, 95% CI 8.22-13.56%) and the number of participants with glycated hemoglobin (HbA1c) level < 7% (RR 2.61, 95% CI 1.29-5.28). Also, HCL significantly reduced the time during which glucoselevel was > 180 mg/dL (WMD-10.46%, 95% CI-13.99 to-6.93%) and the mean levels of glucose (WMD-16.67 mg/dL, 95% CI-22.25 to-11.09 mg/dL) and HbA1c (WMD-0.50%, 95% CI-0.68 to-0.31). There were no significant differences between therapies regarding time during which glucose level was < 70 mg/dL or <54 mg/dL or number of episodes of ketoacidosis, hyperglycemia, and hypoglycemia. In this meta-analysis, HCL compared with control therapies was associated with improved time in range and HbA1c control in children and adolescents with T1D and a similar profile of side effects. These findings support the efficacy of HCL in the treatment of T1D in this population.