The expansion of transcatheter aortic valve implantation (TAVI) to low-risk and younger patients has increased the relevance of the long-term durability of transcatheter heart valves (THV). The present study aims to assess the 10-year durability, hemodynamic performance, and clinical outcomes after TAVI using the CoreValve system. An analysis from a prospective registry with predefined clinical and echocardiographic follow-up included 302 patients who underwent TAVI with the CoreValve system between 2007 and 2015. Bioprosthetic valve failure (BVF) was defined as any bioprosthetic valve dysfunction-related death, re-intervention, or severe hemodynamic valve deterioration. At the time of TAVI, the mean age was 80.41 ± 7.01years, and the Society of Thoracic Surgeons (STS) score was 6.13 ± 5.23%. At latest follow-up (median [IQR]: 5 [2-7] years), cumulative all-cause mortality ratesat 3, 5, 7, and 10years was 23.7%, 40%, 65.8%, and 89.8%, respectively. Mean aortic valve area and transvalvular gradient post-TAVI and at 5, 7, and 10years were 1.94, 1.87, 1.69, and 1.98 cm2 (p = 0.236) and 8.3, 9.0, 8.2, and 10.1mmHg (p = 0.796), respectively. Overall, 11 patients had BVF, of whom six had structural valve deterioration (SVD). The 10-year actual and actuarial freedom from BVF was 96.1% and 78.8%, and from SVD was 97.9% and 80.9%, respectively. Three patients developed significant non-SVD due to severe paravalvular leakage, and two patients were diagnosed with infective endocarditis. Using an early-generation self-expanding bioprosthesis, we documented durable hemodynamic performance and low rates of BVF and SVD up to 10 years after TAVI.