Patients with prior myocardial infarction (MI) have an increased risk of experiencing a secondary event which is exacerbated by mental stress. Our team has developed a miniaturized patch with the capability to capture electrocardiogram (ECG), seismocardiogram (SCG) and photoplethysmogram (PPG) signals which may provide multimodal information to characterize stress responses within the post-MI population in ambulatory settings. As ECG-derived features have been shown to be informative in assessing the risk of MI, a critical first step is to ensure that the patch ECG features agree with gold-standard devices, such as the Biopac. However, this is yet to be done in this population. We, thus, performed a comparative analysis between ECG-derived features (heart rate (HR) and heart rate variability (HRV)) of the patch and Biopac in the context of stress. Our dataset contained post-MI and healthy control subjects who participated in a public speaking challenge. Regression analyses for patch and Biopac HR and HRV features (RMSSD, pNN50, SD1/SD2, and LF/HF) were all significant (p<0.001) and had strong positive correlations (r>0.9). Additionally, Bland-Altman analyses for most features showed tight limits of agreement: 0.999 bpm (HR), 11.341 ms (RMSSD), 0.07% (pNN50), 0.146 ratio difference (SD1/SD2), 0.750 ratio difference (LF/HF).Clinical relevance- This work demonstrates that ECG-derived features obtained from the patch and Biopac are in agreement, suggesting the clinical utility of the patch in deriving quantitative metrics of physiology during stress in post-MI patients. This has the potential to improve post-MI patients' outcomes, but needs to be further evaluated.
Read full abstract