Objective To compare the diagnostic value of digital breast tomosynthesis (DBT) versus digital mammography (DM) in the evaluation mass like lesions in dense breasts. Methods Eight hundred and fifty eight patients, who had DBT and DM, with breast diseases confirmed by pathology were retrospectively reviewed. The breast density and the imaging features were classified according to the standard of American College of Radiology(2013) and breast imaging reporting and data system (BI-RADS) scores. Six hundred and thirty one patients with mass lesions in dense breasts were chosen to included. All 631 patients had DM of both breasts and DBT of affected breast prior to mastectomy. Pathological results were used as the golden standard. Detection rate and diagnostic accuracy of masses, sensitivity and specificity of diagnosis, false negative and recall rates, the margins and spicules and BI-RADS category were calculated. The detection rate and diagnostic accuracy, the margins and spicules of masses were compared with Chi-square test, the BI-RADS scores were analyzed with rank test. Results With DBT, average detection rate increased from 77.3% (488/631) to 84.3% (532/631), average diagnostic accuracy rate increased from 73.4% (463/631) to 82.3% (519/631) ,with statistical significant difference between groups (χ2=27.191,36.890,P<0.01).Sensitivity for DBT and DM were 68.1% (205/301) and 58.8% (177/301) , respectively. Specificity for DBT and DM were 95.2% (314/330) and 86.7% (286/330) , respectively. Recall rate for DBT and DM were 3.6% (23/631) and 9.8% (62/631) , respectively. One hundred and seventy two cases and 75 cases with circumscribed masses of benign cases were detected by DBT and DM, respectively. One hundred and eighty-two cases and 115 cases with spiculated masses of malignant cases were detected by DBT and DM, respectively, with statistical significant difference between groups (χ2=70.471 and 21.707, P<0.01). Of 330 malignant cases, the numbers of BI-RADS category 0, 2, 3, 4A, 4B, 4C, 5 on DM images were 19,4,8,13,52,93,141, respectively, the numbers on DBT images were 3,1,7,5,52,83,179, respectively, with statistical significant difference (Z=-2.235,P=0.025) . Of 301 benign cases, the number of BI-RADS category 0, 2, 3, 4A, 4B, 4C, 5 on DM images were 43,10,60,117,58,11,2, respectively, the number on DBT images were 20,4,64,150,55,7,1, respectively, with no statistical significant difference between groups (Z=-1.846,P=0.065) . Conclusion DBT significantly improves the detection of mass-like lesions in dense breasts, display of mass margin and architectural distortion of surrounding tissue. Key words: Breast neoplasms; Mammography; Digital breast tomosynthesis