Abstract

The purpose of this study was to assess local diagnostic reference levels (LDRLs) for full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) mammography in India. Data from 1500 women were collected from five different mammography facilities in major cities in Tamil Nadu, India. The mean of mean glandular dose were used to arrive at an LDRL. The noted mean compressed breast thickness was 55.26±3.4. The recorded mean MGDs for the five centres were 3.1±0.1 and 3.8±0.2mGy for FFDM and DBT, respectively. The 75th percentile value for all five centers is 3.3 and 4.0mGy for FFDM and DBT, respectively. The LDRLs found in the current study were also compared with those from earlier studies conducted in other nations, such as the United Kingdom, Malaysia, Morocco, and Ghana. The present study is the first of its kind to determine the LDRL for the FFDM and DBT scanners operating in the Tamil Nadu region, India, and is proposed as a starting point that will allow professionals to evaluate and optimize their practice. Furthermore, similar studies in other regions of India are necessary in order to establish National DRLs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.