Small diameter (mini) dental implants have become more popular in recent years as alternatives to classical implant treatment in clinical cases with critical bony situations. However, an in-depth scientific analysis of the mechanical and biomechanical effects of small diameter implants has not yet been published. The aim of the present study was to investigate experimentally different commercial mini implants by measuring their displacements under immediate loading. Twelve commercially available mini implants were measured. Implants were inserted into porcine mandibular segments and loaded by means of a predefined displacement of 0.5 mm of the loading system. The implants were loaded at an angle of 30° to the implant long axis using the self-developed biomechanical hexapod measurement system. Implant displacements were registered. The experimental results were compared to the numerical ones from a previous study. Measured implant displacements were within the range of 39-194 μm. A large variation in the displacements was obtained among the different implant systems due to the different designs and thread profiles. Comparing experimental and numerical results, the displacements that were obtained numerically were within the range of 79-347 μm. The different commercial mini implants showed acceptable primary stability and could be loaded immediately after their insertion.