The pig (Sus scrofa) is the most widely used large animal model in Europe, with cardiovascular research being one of the main areas of application. Adequate refinement of interventional studies in this field, meeting the requirements of Russell and Burch's 3 R concept, can only be performed if blood-contacting medical devices are hemocompatible. Because most medical devices for cardiovascular interventional procedures are developed for humans, they are tested only for compatibility with human blood. The aim of this study was therefore to determine whether there are differences in behavior of human and porcine platelets from commercial hybrid pigs when they come into contact with borosilicate glass, which was used as an exemplary thrombogenic material. For this purpose, changes in platelet count, platelet volume and platelet expression of the activation markers CD61, CD62P and CD63 were measured using a modified chandler loop-system simulating the fluidic effects of the bloodflow. Commercial hybrid pig and human platelets showed significant adhesions to borosilicate glass but the commercial hybrid pigs platelets showed a significantly higher tendency to adhere to borosilicate glass. In contrast to human platelets the platelets of commercial hybrid pigs showed significant activation after 4 to 8 minutes exposure to borosilicate glass and there were differences among the ratios of surface and activation markers in between the platelets of both species.