BackgroundSequential modifications to the manufacturing process of highly cross-linked polyethylene (HXLPE) have improved the wear resistance and implant survival of these liners in total hip arthroplasty (THA). However, no study has examined the long-term (mean 10 year) wear rates and clinical outcomes of third-generation HXLPE in THA. The aim of our study was to report the longest-to-date analysis of wear rates and clinical outcomes of a third-generation HXLPE liner. MethodsA series of 133 THAs using a specific HXLPE acetabular liner were retrospectively evaluated. Linear and volumetric wear rates were determined using a validated radiographic technique and clinical outcomes were analyzed. Multivariate analyses were performed to determine risk factors for accelerated wear. ResultsAt a mean follow-up of 10.4 years (range, 8 to 13.4), the mean linear wear rate was 0.0172 mm/year and the mean volumetric wear rate was 16.99 mm3/year. There were no instances of osteolysis or mechanical failures at any time point and there was a 100% acetabular component survival rate. Younger age and use of offset liners were independent risk factors for increased wear (P < .01 for both). ConclusionsOur series of a third-generation HXLPE demonstrated very low wear rates and excellent implant survival at a mean of 10.4 years following primary THA. Future comparative studies at the 15- and 20-year follow-up timepoints are necessary to determine if such findings translate to true improvements in the tribological properties and longevity of these liners when compared to previous generations of HXLPE liners.