Frictional torque of VEXLPE and UHMWPE acetabular liners against CoCr and ZTA femoral heads was measured with a multidirectional hip joint simulator for a wide range of diameters. Measured frictional torque varied from 1 Nm to 7 Nm. The contact temperature was estimated with a subsurface thermocouple within the liner, located near the contact surface. When the estimated contact temperature was 37 °C, the lubricant bulk temperature was 30.1 ± 1.1 °C. When the lubricant bulk temperature was 37 °C, the estimated contact temperature was 43.4 ± 1.1 °C. The custom to use 37 °C as the lubricant bulk temperature in wear testing may lead to overheating of the contact and unrealistically low wear rates, as indicated by recent studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.